BetaFLEX
CALIBRATOR
|
-
Universal I/O through
plug-in modules for single and dual pressure (±0.025% accuracy),
thermocouples, RTDs, frequency, and dc current and voltage
-
BetaSmart option enables
calibration/configuration of Honeywell digitally enhanced (DE) protocol
Smartline transmitters
-
Fully compatible with
Hathaway's MasterCAL Calibration System Software and Honeywell's DocuMint
documenting software
-
Input modules feature
measurement, simulation, and autoranging
|
|
Hathaway's
BetaFLEX portable documenting calibrator uses plug-in I/O modules that
give it the industry's most extensive calibration capabilities. It can
calibrate just about any type of process instrument a field technician
is likely to encounter.
Exclusively, through an
optional BetaSmart enhancement and use of a Honeywell SCT 3000 PCMCIA card,
BetaFLEX can even automatically calibrate, document, and configure Honeywell
digitally enhanced (DE) protocol Smartline transmitters (specifically,
ST 3000 Smart Pressure, STT 3000 Smart Temperature, and SMV 3000 Smart
Multivariable models). In addition to automated operations, it can manually
read and write data, perform routine calibrations, and make operating adjustments
to the Honeywell transmitters, all functions that previously required a
dedicated handheld configurator/communicator or portable computer.
The BetaSmart option allows
the user to adjust the transmitter zero, span, URV, LRV, and curve type
(where applicable), to test the internal ND and DIA converters, and to
read and/or write data on the transmitter's scratch pad data area. The
transmitter's diagnostic status can also be determined and acted upon. |
|
For automated calibration operations,
BetaFLEX is fully compatible with Hathaway's MasterCAL Calibration System
Software and Honeywell's DocuMint documenting software. Other third party
software may be available as well.
BetaFLEX works with all
of Hathaway's BetaPort calibration modules. BetaPort-P pressure range modules
have industry standard ±0.025% accuracy and can measure gauge, vacuum,
absolute, and differential pressure (23 ranges are available from 5 psi/330
mBar up to 10,000 psi/670 Bar full scale). Other modules include the BetaPort-D2
dc current and voltage module, the BetaPort-R multicurve RJD and resistance
module, the BetaPort-T multicurve thermocouple module, and the BetaPort-F
frequency module. Modules include both measurement and simulation functions
and are autoranging.
Users of Hathaway's BetaGauge
II
digital pressure gauge, which also employs BetaPort pressure and dc current/voltage
calibration modules, can have their units upgraded with BetaFLEX and BetaSmart
capabilities at Hathaway's Carrollton, TX manufacturing facility. |
Available
Ranges and Accuracies |
-
BetaPort-P Pressure
Modules
Gauge (psig)
|
Gauge (psig)
|
Vacuum (psig)
|
Absolute (psia)
|
Compound
|
Differential (psid)1
|
0-5 (0-350 mBar)4
|
0-300 (0-20 Bar)
|
0 la –5 (0 la –350 mBar)4
|
0-15 (0-1 Bar)3
|
-15 la 15 (-1 la 1 Bar)3
|
0-5 (0-350 mBar)4
|
0-7,2 (0-500 mBar) 2,3
|
0-1000 (0-70 Bar)
|
0 la –15 (0 la –1 Bar)3
|
0-30 (0-2 Bar)
|
-15 la 30 (-1 la 2 Bar)
|
0-30
(0-2 Bar) |
0-10
(0-700 mBar) 3 |
0-1500
(0-100 Bar) |
-
|
0-50
(0-3,5 Bar) |
-
|
-
|
0-30 (0-2 Bar)
|
0-3000 (0-200 Bar)
|
-
|
0-100 (0-7 Bar)
|
-
|
-
|
0-50 (0-3,5 Bar)
|
0-5000 (0-340 Bar)
|
-
|
-
|
-
|
-
|
0-100 (0-7 Bar)
|
-
|
-
|
-
|
-
|
-
|
0-150 (0-10 Bar)
|
-
|
-
|
-
|
-
|
-
|
|
1
Maximum static pressure, 200 psig (14 Bar)
2200
InWc
3
Thermal and Pressure Hysteresis: ,0025 psi (,1724 mBar)
4
Thermal and Pressure Hysteresisv: ,0030 psi
(,2068 mBar)
All other ranges: No Hysteresis
Other Pressure Units (keypad
selectable): Kp/cm2; Atmos; mBar; Bar; Kpa; mmHg; InWc 4°C, 20°C,
60°F; plus one user-defined unit; choice of engineering unit may be
restricted by available resolution.
ACCURACY
Gauge, Vacuum, Absolute,
Compound, and Differential Standard Ranges:±0.025% FS (0°C to
50°C) for six months
Gauge Extended Ranges:±0.035%
FS (0°C to 50°C) for six months
Accuracy for ranges with
thermal and pressure hysteresis is: ±(Accuracy%FS + Hysteresis in
psi) |
|
-
BetaPort-D2
do current and voltage module
Description: General
purpose dc measurement and simulation by sourcing and measuring dc voltage
and current. Output mode can sink and source in both voltage and current
functions. When sinking voltage, unit will ramp up to 21 mA. When in mA
dc output mode, unit will detect and signal an open loop condition. |
Function
|
Range
|
Accuracy *
|
Resolution
|
dc Volts In |
0 la 110 mV
|
± 0,01% + 1 LSD
|
0,01 mV dc
|
0 la 5,1 mV
|
± 0,01% + 1 LSD
|
0,0001 Vdc
|
0 la 12,1 mV
|
± 0,01% + 1 LSD
|
0,001 Vdc
|
0 la 25 mV
|
± 0,025% + 1
LSD
|
0,01 Vdc
|
dc Volts Out |
0 la 110 mV
|
± 0,01% + 1 LSD
|
0,01 mV dc
|
0 la 5,1 mV
|
± 0,01% + 1 LSD
|
0,0001 Vdc
|
0 la 12,1 mV
|
± 0,01% + 1 LSD
|
0,001 Vdc
|
dc mA In |
0 la 20,5 mA
|
± 0,01% + 1 LSD
|
0,001 mA dc
|
0 la 51 mA
|
± 0,025% + 1
LSD
|
0,1 mA dc
|
dc mA Out ** |
0 la 20,5 mA
|
± 0,015% + 1
LSD
|
0,001 mA dc
|
* ln percent of fullscale
** Can drive 20.5 mA dc through a loop resistance of 600 ohmi minimum
All ranges can be displayed
as a percent ot scale based on a nominal range as described here: |
Range
|
Nominal Range
|
% Display
|
0 la 110 mV
|
0 la 100 mV
|
0 la 110 %
|
0 la 5,1 V dc
|
1 la 5 Vdc
|
(-)25 la 102%
|
0 la 12,1 Vdc
|
0 la 10 Vdc
|
0 la 120%
|
0 la 25 Vdc
|
0 la 25 Vdc
|
0 la 100%
|
0 la 20,5 mA
|
4 la 20 mA dc
|
(-)25 la 103%
|
0 la 51 mA
|
10 la 50 mA dc
|
(-)25 la 102%
|
|
|
|
-
BetaPort-F Frequency
Module
|
Description: Measurement
and sourcing of frequency signals. Output amplitude adjustable between
0.05 and 12 Volts peak-to-peak. Input mode operates with peak-to-peak voltage
of 0.05 to 240 Volts. Input mode adjustable threshold of 0.01 to 1.0 Volts
for noise rejection. Module provides bipolar sine and square wave outputs
and zero based square wave outputs. Input mode reads both bipolar and zero-based
signals.
Accuracy: ±0.01
% of reading for all ranges in both measure and simulate modes |
Range
(Measure and Simulate)
|
Unit
|
Resolution
|
100,00
|
kHz
|
1 Hz
|
10,00
|
kHz
|
0,1 Hz
|
1000,00
|
Hz
|
0,01 Hz
|
100,00
|
Hz
|
0,01 Hz
|
1000,00
|
CPM *
|
0,1 CPM
|
100,00
|
CPH **
|
0,01 CPH
|
* Counts Per Minute
** Counts Per Hour
|
|
|
-
BetaPort-R Multicurve
RTD Module
|
|
Description:
Measurement and simulation of common RID signals in 2, 3, or 4 wire modes
(all specifications here are for 3 wire mode). Simulation circuit can detect
pulsed excitation currents commonly found in "smart" RTD transmitters.
Can use an excitation current as low as 0.1 mA.
Accuracy: For simulation
of output functions, ±0.025% of resistance span for all ranges.
For measurement of all input functions, ±0.025% of resistance span
for all temperature ranges; ±0.5ohms for measurement of ohms. |
|
Function
|
Alpha
|
Range
|
Resolution
|
Pt100 (385) Simulate
|
1,3850
|
-200 la +850°C
|
0,1°C
|
Pt100 (385) Measure
|
1,3850
|
-200 la +850°C
|
0,1°C
|
Pt100 (392) Simulate
|
1,3916
|
-200 la +649°C
|
0,1°C
|
Pt100 (392) Measure
|
1,3916
|
-200 la +649°C
|
0,1°C
|
Pt100 (390) Simulate
|
1,3902
|
-195 la +649°C
|
0,1°C
|
Pt100 (390) Measure
|
1,3902
|
-195 la +649°C
|
0,1°C
|
Pt100 (392) Simulate
|
1,3926
|
-200 la +862°C
|
0,1°C
|
Pt100 (392) Measure
|
1,3926
|
-200 la +862°C
|
0,1°C
|
Cu 10 Simulate
|
1,4272
|
-200 la +260°C
|
1,0°C
|
Cu 10 Measure
|
1,4272
|
-200 la +260°C
|
1,0°C
|
Ni 120 Simulate
|
1,6720
|
-80 la +320°C
|
0,1°C
|
Ni 120 Measure
|
1,6720
|
-80 la +320°C
|
0,1°C
|
Ni 110 Simulate
|
1,5801
|
-105 la +310°C
|
0,1°C
|
Ni 110 Measure
|
1,5801
|
-105 la +310°C
|
0,1°C
|
Ohms Simulate
|
-
|
0 la 1.000 ohms
|
0,01 ohms
|
Ohms Measure
|
-
|
0 la 1.000 ohms
|
0,01 ohms
|
|
|
|
-
BetaPort-T Multicurve
Thermocouple Module
Description: Measurement
and sourcing of common thermocouple signals. In output mode, module will
have a minimum drive capacity of 8 mA. In measurement mode, unit will detect
and indicate an open (burnout) condition.
Accuracy: Based on
a dc measurement and simulation range accuracy of ±0.01 % of the
millivolt range of any curve with a minimum span of 10 mV, and a cold junction
measurement accuracy of +0.1°C within the ambient temperature range
of 10 to 40°C. Outside that range, the effect of ambient temperature
change on cold junction temperature measurement will be less than 0.005%
per °C |
Function
|
Range
|
Resolution
|
Type J (ANSI)
Measure or Simulate |
-210 la –180,0°C
-179,9 la –50,0°C
-49,9 la +500,0°C
500,1 la 1200,0°C
|
0,1°C
0,1°C
0,1°C
0,1°C
|
Type K (ANSI)
Measure or Simulate |
-230,0 la –100,0°C
-99,9 la +1050,0°C
1050,1 la 1372,0°C
|
0,1°C
0,1°C
0,1°C
|
Type T (ANSI)
Measure or Simulate |
-260,0 la –200,0°C
-199,9 la –50,0°C
-49,9 la 0,0°C
0,1 la 400,0°C
|
0,1°C
0,1°C
0,1°C
0,1°C
|
Type E (ANSI)
Measure or Simulate |
-240,0 la –200,0°C
-199,9 la –100,0°C
-99,9 la +850,0°C
850,0 la 1000,0°C
|
0,1°C
0,1°C
0,1°C
0,1°C
|
Type R/S (ANSI)
Measure or Simulate |
-50,0 la 299,0°C
300,0 la 1768,0°C
|
1,0°C
0,1°C
|
Type B (ANSI)
Measure or Simulate |
100,0 la 199,0°C
200,0 la 499,0°C
500,0 la 999,0°C
1000,0 la 1820,0°C
|
1,0°C
1,0°C
1,0°C
0,1°C
|
Type N (ANSI)
Measure or Simulate |
-250,0 la –226,0°C
-225,0 la –101,0°C
-100,0 la +1300,0°C
|
1,0°C
1,0°C
0,1°C
|
Type G (Nota 1) Measure
or Simulate |
150,0 la 2310,0°C
|
1,0°C
|
Type C (Nota 2) Measure
or Simulate |
50,0 la 2310,0°C
|
1,0°C
|
Type D (Nota 3) Measure
or Simulate |
50,0 la 2310,0°C
|
1,0°C
|
Type L (DIN)
Measure or Simulate |
-200,0 la –151,0°C
-150,0 la +900,0°C
|
0,1°C
0,1°C
|
Type T (DIN)
Measure or Simulate |
-200,0 la –101,0°C
-100,0 la +600,0°C
|
0,1°C
0,1°C
|
Type M Measure or Simulate |
0 la 1310,0°C
|
1,0°C
|
Platinel |
0 la 1115,0°C
1116,0 la 1400,0°C
|
1,0°C
1,0°C
|
mV Measure or Simulate |
-100,00 la +100,00
|
0,01 mV
|
|
|
Note 1: Tungsten
vs. Tungsten-26% Rhenium
Note 2: Tungsten-5%
vs. Tungsten-26% Rhenium
Note 3: Tungsten-3%
Rhenium vs. Tungsten-26% Rhenium
One user definable curve
of 32 segments is also provided. |
|
SPECIFICATIONS |
ENVIRONMENTAL |
Temperature:
No effect from 15 to 35°C, +0.005%/°C or better outside that
range
Relative Humidity:
No effect, 0 to 95% non-condensing
Overpressure (without
damage or calibration effects)
a. 0-15 Psi* 400% of sensor
range
b. 30-250 psi 200% of sensor
range
c. 300-10,000 psi 150% of
sensor range
*0-200 InWc is 300% of sensor
range
OTHER
Alarm Trip Detect:
Wet contact input, 48 Vdc. max.
Damping Factor: Programmable
1.0 to 5 sec response time for all inputs (running average)
Display: 128 X 128
pixel LCD with LED backlight; two inputs simultaneously displayed with
maximum of 5 digits resolution
POWER
Power Supply: Rechargeable
NiCad external intrinsically safe battery pack |
|
Pressure Connections:
1/8
“ FNPT 316SS
Electrical Connections:
jack standard de ¾” central
Temperature Pt1OOI39O
Burns, Connections: 7 RTD types (Ptl 00/385D1N, Ptl 00/392J IS, PtlOO/390
Burns, PtlOO/392US, CulO, Nil 10, and Ni120); round and spade-type miniature
TIC plugs, as well as bare wire |
MECHANICAL
Weight: <1,5 kg
with carrying case and accesories (typical)
Size: 4.3" x 7.5"
x 2.4"
(110 mm x 190 mm x 60 mm) |
|
|
|
|
|
|
|